What is medical suction measured in? Two independent quantities: the vacuum level, given in kPa, mmHg or mbar, and the flow, given in litres per minute of free air. A buyer who specifies only one of them risks a unit that looks strong on paper but fails at the bedside. This explainer shows how to read both figures, how to convert between units and how to match settings to the clinical task.
Vacuum: suction pressure in kPa, mmHg and mbar
Suction pressure is a negative pressure measured relative to the atmosphere. It appears in several units depending on country and manufacturer: kilopascals (kPa), millimetres of mercury (mmHg), millibar (mbar) and bar. The useful conversions are 1 kPa = 10 mbar = 7.5 mmHg, and 100 kPa = 1 bar = about 750 mmHg.
A reading is usually printed with a minus sign, such as −98 kPa. For example, the ATMOS Record 55 lists a maximum vacuum of −98 kPa (−980 mbar), and the ATMOS C 451 lists −91 kPa (−910 mbar). When a supplier quotes a suction pressure kPa figure and another quotes suction pressure mmHg, convert them before comparing.
- 1 kPa = 10 mbar = 7.5 mmHg
- 100 kPa = 1 bar = approximately 750 mmHg
- −90 kPa corresponds to about −675 mmHg
Flow: suction flow rate in l/min
Flow tells you how much air the pump can move, usually quoted as free air at the pump inlet. It determines how fast the unit clears fluid and how quickly vacuum recovers when the tubing opens. The ATMOS Record 55 is rated at 55 l/min and the ATMOS C 451 at 45 l/min.
A unit with high maximum vacuum but low flow will reach a high gauge reading when the tubing is clamped, yet clear a bleeding wound slowly. That is why a high vacuum high flow suction specification is the typical requirement for operating rooms, while drainage applications require the opposite.
Typical settings by application
Required vacuum depends on the application and is set by clinical protocol. In general, airway suction and OR suction use moderate to high vacuum with high flow; gastric suction often uses low, sometimes intermittent vacuum; chest drainage uses very low vacuum, measured in millibar; and neonatal suction uses reduced vacuum. See our guide on neonatal suction for the special case of newborns.
The maximum a unit can produce is not the setting you use. A good unit lets you set the working vacuum accurately and limits what can reach the patient, often through a regulator; see medical suction regulators. For the broader context of device types, read types of medical suction.
How standards classify suction equipment
ISO 10079 is the family of standards for medical suction equipment, and it classifies equipment by vacuum and flow as well as by power source. Equipment is typically described as high vacuum/high flow, high vacuum/low flow or low vacuum/low flow, with overflow protection and a bacterial filter among the safety requirements. Check the current edition of the standard and ask the supplier for the declaration of conformity; our article on ISO 10079 explains the points.
Neither figure alone describes performance. Ask the supplier for a curve or table showing flow at several vacuum levels, and for the test conditions, including tubing length and canister size.
How to specify suction when buying
Write the specification as a pair of numbers for each use: for example, minimum vacuum at the gauge, minimum flow at the working vacuum, noise limit and canister volume. Include the operating mode (continuous or intermittent), the mains frequency, and whether battery operation is required.
Compare real devices on the same basis, for instance the ATMOS Record 55 against the ATMOS C 451, and test with your own tubing. Confirm UAE registration and contact Beckfinity for a quote or consultation on specifying your suction units.
Frequently asked questions
How do I convert kPa to mmHg for suction?
Multiply kPa by 7.5 to get mmHg. For example, 90 kPa is about 675 mmHg. One kPa equals 10 mbar.
Which is more important, vacuum or flow?
Both. Vacuum shows how strongly the pump can pull when blocked; flow shows how quickly it moves fluid and air when open. The task decides the balance.
What does high vacuum high flow mean?
It describes suction equipment capable of both a high vacuum level and a high air flow, as is needed in operating rooms and for airway suction. Check the definition in the current edition of ISO 10079.
Is the maximum vacuum the setting I should use?
No. The working vacuum is set by clinical protocol and is usually well below the maximum the unit can produce.
Related articles
- Portable Suction Machine Guide: Flow, Vacuum, Canisters and Battery
- Portable Suction Machine for Home Use: How Providers and Clinics Choose
- ENT Treatment Unit Guide: Functions, Specifications and Selection
- Surgical Smoke Evacuation Systems: Filtration, Accessories and Selection Criteria
- Medical Suction Machine Price and Procurement in the UAE: A Buyer's Guide
- Surgical Smoke Safety: Plume Hazards and Evacuation Guidelines for the OR
- How to Clean a Portable Suction Machine: Disinfection, Filters and Troubleshooting
- Surgical Suction Device Guide: Flow, Vacuum, Canisters and Noise
- Medical Suction Canisters, Liners, Tubing and Filters: A Compatibility Guide
- Choosing a Tissue Retrieval Bag for Gynecologic Laparoscopy
- Skin graft mesher guide for burn and plastic surgery units
- Care and Maintenance of Surgical Instruments: A Practical OR and CSSD Guide